Instrument type | Automatic Chemiluminescent Immunoassay Analyzer |
Model | POClia 8 |
Dimensions | 530(W)*626(D)*426(H)nm |
Weight | 30kg |
Test batch | 8-channel parallel detection |
Testing time | Depending on the reagent item |
Sample | Whole blood/Serum/Plasma/Urine (depends on reagent type) |
Principle | CLEIA and Magnetic particle separation technology |
Sampling accuracy | CV<1%,100ul for example |
Temp control of reaction area | 42 ± 0.5℃ Fluctuation:≤1.0℃ |
Spectra module of test module | 300nm ~ 650nm |
Linear correlation | Within a concentration range of not less than 2 orders of magnitude, the linear correlation coefficient (r) ≥ 0.99 |
Inter-batch repeatability | (CV,%)≤8% |
Stability | Results CV at 4 hour, 8 hour with the initial time when machine is steady is ± 10% |
Data storage | Patient date: >100000 |
QC Date | >100000 |
Calibration Date | >100000 |
Power supply | 100-240 VAC , 50Hz , 200VA |
Printer | Built-in |
Computer | Built-in |
Screen | LCD touch-screen |
24h operation (standby) | Recommend |